Digitizing Oscilloscope

                 MODEL  :  GDS-830

        Digital Storage Oscilloscope  100MHz

 

SPECIFICATIONS

Display System

CRT Type

7-inch diagonal, magnetic deflection; horizontal raster scan; green phosphor

Video Display Resolution 480 (Vertical) x 640 (Horizontally) dots
Waveform Display Graticule 8 x 10 divisions
Intensity Levels Adjustable overall intensity and contrast

Vertical System

CH1 and CH 2 2mV / DIV ~ 5V / DIV
Accuracy +3%
Bandwidth DC ~ 100MHz (-3dB)
Rise Time 3.5nS approx.
Input Coupling AC, DC, GND
Input Impedance 1M£[ + 1%, ~ 22pF
Maximum Input 300V CAT II (DC + AC peak)
Polarity Normal & Invert
Waveform Processing CH1-CH2, CH1+CH2, invert
Offset Range 2mV/DIV ~ 50mV/DIV :  + 0.5V, 100mV/DIV ~ 500mV/DIV : + 5V, 1V/DIV ~ 5V/DIV : + 50V
Bandwidth Limit 20MHz (-3dB)

Horizontal

Range 2nS/DIV ~ 5S / DIV (1-2-5 increments)
Modes MAIN, WINDOW, DELAY, ROLL, X-Y
Accuracy +0.01%
Delay Range Pre-trigger : 10 DIV, Post-trigger : 500 DIV

Signal Acquisition System

Sampling Rate 100MSa/s
Vertical Resolution 8 bits
Record Length / Channel 125k dots
Single Shot Record Length 125k dots
Single Bandwidth 10MHz
Peak Detection 20nS (10µS / DIV ~ 5V / DIV)
Average 2, 4, 8, 16, ..., 256

Trigger

Source CH1, CH2, LINE, EXT
Modes AUTO-LEVEL, AUTO, NORMAL, SINGLE, TV, TIME-DELAY, EVENT-DELAY
Coupling AC, DC, LF rej, HF rej, Noise rej
Sensitivity DC ~ 25MHz : approx. 0.35 DIV or 3.5mV ;
25MHz ~ 100MHz : approx. 1 DIV or 10mV
TV TV trigger sensitivity : 0.5 division of sync signal

X-Y Mode

X-Axis Input/Y-Axis Input Channel 1 (CH1) / Channel 2 (CH2)
Bandwidth As vertical system
Phase Shift +3° at 100kHz

Cursor & Measurement

Automated Voltage Measurement Vhi, Vlo, Vmax, Vmin, Vpp, Vavg, Vrms
Automated Time Measurement Rise Time, Fall Time, Duty Cycle, Frequency, Period, Positive Width, Negative Width
Cursor Measurement Voltage difference between cursors (£GV)/time difference between cursors (£GT)

Control Panel Functions Autoset "Autoset" adjust vertical volt/DIV, horizontal sec/DIV, and trigger level automatically
Save/Recall Up to 15 sets of measurement conditions can be saved and recalled
Waveform Trace Save/Recall 2 sets of waveform can be saved and recalled

External Trigger Range ±12V
Sensitivity DC ~ 50MHz : ~ 50mV,
50MHz ~ 100MHz : ~ 100mV
Input Impedance 1M£[ ±1%, ~ 22pF
Maximum Input 300V CAT II (DC + AC peak)

Miscellaneous Parallel port A 25-pin IBM PC type, parallel printer interface printer compatibility CANON, EPSON, DeskJet, LaserJet. more info
RS-232 Interface A 9-pin male DTE RS-232 interface that complies with EIA  / TIA 574-90
GPIB Interface Option; fully programmable with IEEE 488.2 compliance
Video Signal Output DB-15 rear panel video connector, VGA compatible.

Power Source Line Voltage Range 100V ~ 240V AC, automatic selection
Line Frequency 48Hz ~ 63Hz
Power Consumption 85Watts, 95VA maximum, with fan

Accessories Power cord x 1 ;
Instruction manual x 1 ;
Probes(10:1/1:1) x 2

Dimensions & Weight 330(W) x 155(H) x 385(D) mm, Approx. 7 kg